Agilent 1100 Series (G1314A) UV-Visible HPLC System

Used Lab Equipment Warranty Condition: Used Will be thoroughly tested and serviced
6 MONTH IN SHOP WARRANTY
Stock #:
36607A
Status:
1 Available In Stock (request a quote)
Dimensions (approx.):
14" x 17" x 7" (outside)
Electrical:
100-240 Volts,
Product Description Consisting of the following components:
Mobile Phase tray
G1322A; Degasser
G1314A; Variable Wavelength Detector (VWD)
G1316A; TCC (thermal column compartment)
G1313A; ALS
G1311A; Quaternary Pump
COMPUTER: Hardware: P4 2.80 GHz, 504 MB RAM, 80gig IDE hard drive with C: and D: partitions. Two NICs.
OS: WinXP SP3
Software: Chemstation for GC, LC, CE, A/D, CE/MSD & LC/MSD. Rev A10.02, 2004

The G1314A Variable Wavelength Detector lowers detection limits with simplified optics for high sensitivity, low baseline drift, and wide linear range. This wide linear range allows for reliable simultaneous quantitation of main and side products and impurities. The Agilent 1100 VWD's minimum baseline noise gives it the highest baseline stability for the lowest detection limits. This VWD offers programmable wavelength switching for sensitivity and optimized selectivity optimized to each analyte's elution, along with stop-flow wavelength scanning for quick wavelength optimization.
SPECIFICATIONS:
- Detection type: Double-beam photometer
- Light Source: Deuterium lamp
- Wavelength Range: 190 - 600 nm
- Short term noise (ASTM) Single and Multi-Wavelength: ±0.75 × 10-5AU at 254 nm
- Drift: 3 × 10-4AU/hr at 254 nm
- Linear Absorbance range: > 2 AU (5%) upper limit
- Wavelength accuracy: ± 1 nm
- Slit width: 1, 2, 4 , 8, 16 nm
- Band Width: 6.5 nm typical
- Diode Width: < 1 nm

The G1313A ALS holds a 100 position Sample Tray and is a component in Hp Agilent 1100 series stackable HPLC system.
SPECIFICATIONS:
- GPIB (IEEE-448), Serial (RS232C)
- GLP features: Early maintenance feedback (EMF), records of maintenance and errors
- Safety features: Leak detection, safe leak handling, and error detection and display.
- Replicate injections: From 1-99 from one vial
- Injection cycle time: Typically 50 sec, depending on draw speed and volume
- Sample capacity: 100 × 2-mL vials in a standard tray; 40 × 2-mL vials in a standard tray; 15 × 6-mL vials in a half tray; Microvials (100 or 300 µL) with sleeves
- Reliable injections from 0.1 µL to 100 µL
- Easy adaption for injection volumes up to 1500 µL for applications ranging from microbore to semipreparative chromatography
- Low internal volume of 300 µL for the minimum contribution to a system's total internal volume
- Overlapped injections for increased productivity

The G1316A COLCOM is a 1100 Series G1316A thermostatted column compartment is built with two heat exchangers for solvent pre-heating, automated column identification module and CAN cable.
SPECIFICATIONS:
- Temp Range: 10° below ambient to 80° C
- Temp Stability: ±0.15°C
- Temp Accuracy: ±0.8°C
- Column Capacity: Three 30cm columns
- Warm-up/Cool Down: 5 mins from ambient to 40°C
- Differant Types Communication Ports: CAN, GPIB, RS-232C, APG Remote

The G1311A Quaternary Pump is an optimized design ensuring virtually pulse-free and stable solvent flow with dual floating pistions. An inlet valve electronically activated and synchronized to piston stroke eliminates vapor formation with volitale solvents and is maintenance free. Instrument comes equipped with four solvent inlets with built-in gradient control. The gradient formation is performed by a low pressure gradient valve.
SPECIFICATIONS:
- Setable Flow Range: 0.001 - 10ml/min - in 0.001ml/min increments
- Flow Range: 0.2 - 10.0 ml/min
- Recommended pH Range: 1.0 - 12.5, solvents with pH > 2.3 should not contain acids which attack stainless steel
- Gradient Formation: Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve. Delay volume 800-1100µl, depending on back pressure