Primus PSS5-A-MESD Floor-model Autoclave Sterilizer

Used Lab Equipment Warranty Condition: Used Will be thoroughly tested and serviced
6 MONTH IN SHOP WARRANTY
Stock #:
36518A
Status:
2 Available In Stock (request a quote)
Dimensions (approx.):
20"W x 20"D x 38"H (inside)
34"W x 44"D x 74"H (outside)
Electrical:
120 Volts, 60 Hz, 10A
Product Description This 2010 Primus stainless steel single door sterilizer includes Sussman ES-24 boiler as well as autoclave loading cart. Also includes vacuum, gravity, liquids, and test/vac cycles. For surgical packs, select one of the vacuum cycles. For the sterilization of liquids in open or vented containers, select a liquid cycle. The sliding door design reduces the high service cost associated with conventional hinged arm doors and provides an added margin of operator safety. PRIMUS vertically operating doors operate with minimal hand pressure.
A PSS500 microcomputer will offer a selection of eight separate programs for the sterilization of wrapped goods, hard-goods, liquids and test cycles. Cycle parameters may be set by the user and are adaptable to a wide range of products. Alarms are annunciated, displayed and printed. Cycle process parameters will be displayed and recorded throughout the cycle. A side-mounted control panel and a thermal printer dot-matrix technology, with no moving parts in the print head, and 32 characters per line printing. A copy of the sterilization cycle will be available at the end of the run.

Factory Preset Cycles (PSS500 Control):
Double wrapped instruments
Unwrapped nonporous single instrument
Double wrapped instrument trays
Textile Pack
Unwrapped nonporous
single instrument
Fabric packs maximum size: 12x12x20
Vented Borosilicate glass containers, 500ml or smaller
Bowie-Dick Test

PRIMUS sterilizers are designed and manufactured in the USA at a quality controlled ASME facility according to Quality Management Systems which are in compliance with ISO 9001, ISO 13485, Canadian Medical Device Regulation (CMDR) and FDA's Good Manufacturing Practice (GMP) for Medical Devices: General Regulation (21CFR Part 820).

Includes Sussman ES-24 boiler, which is a 3-phase unit. Suitable for industries requiring efficient, high quality saturated steam, including laboratories, hospitals, industrial plants, pilot plants, food and chemical processing applications. Perfect for aImost any application requiring steam from 0-100 PSIG. Each unit is complete with automatic features, including pressure and liquid level controllers, ASME Code pressure vessel and safety valve and are fully tested under power.
KW Rating: 24
BHP rating: 2.45
Lbs/hour steam (at 212°F with 50°F feedwater): 72.3
Voltage: 208
Total amp draw: 67
Size: 23 x 33 x 44"H

ES-STANDARD FEATURES:

WATER FEED SYSTEM: Strainer, solenoid valve and check valve factory wired and plumbed.
LOW WATER CUT-OFF/LEVEL CONTROL: McDonnell Miller No.150 control automatically maintains proper water level, shutting off the boiler when water supply in the boiler drops below a safe operating level.
WATER LEVEL SIGHT GLASS: Allows constant observation of water level while boiler is in operation.
MAIN ON/OFF SWITCH: Allows manual operation of the boiler operating control circuit.
PILOT LIGHT: Indicates control circuit's on/off condition.
INTEGRAL POWER CONTACTORS: Magnetic contactors for energizing the boiler elements. Integrally mounted in the control unit
BLOWDOWN/DRAIN VALVES: Facilitates emptying the boiler pressure vessel and MM150 water column piping during blowdown sequence.
LONG LIFE HEATING ELEMENTS: Industrial grade, heavy duty 0.420 inch diameter stainless steel heating elements,equipped with one piece resistance welded terminations for added strength and safety.
OPERATING PRESSURE CONTROL: Resets automatically to maintain preset pressure within boiler.
ENERGY SAVINGS AND MINIMUM MAINTENANCE: Pressure vessel insulation minimizes heat loss and maximizes energy savings, insured by fibrous glass material.
EASY CONTROL MAINTENANCE: All control panels and components are easily accessible. Fully-louvered openings avoid component heat build-up.
NEMA 1 Louvered enclosure standard


The following link provides additional information from the instruction manual or original manufacturer's brochure. Images, specifications and description refer to the item as it was when new. Right-click to open the PDF in a new window or download it:
http://www.labequip.com/stock/pictures/36518a.pdf